GX-2003 Personal Sample 5-in-1 Gas
Monitor
|
|
|
Features
-
Compact size, weighs only 11 ounces
-
Monitors Combustibles, O2, CO, and H2S
-
0 to
100% volume Methane, Isobutane, and Hydrogen
Auto-ranging display of % volume and % LEL
-
Internal sample drawing pump with 50 foot range
-
Vibration, visual, and audible alarm types
-
Automatic backlight during alarms
-
Calibration reminder and lock out
Ni-Cad or alkaline power source
Battery pack charges in or out of the instrument
-
Quick
charge (complete charge in 90 minutes)
-
Glove
friendly glow in the dark buttons
-
Alarm
latching or non latching
High impact rubber boot protection
-
Up to
600 hours of datalogging with alarm trends
-
Logs
last 20 calibration events
Autocalibration or single calibration
-
TWA
and STEL readings with lunch-break mode
-
Intrinsically safe, CSA, C / US approved
-
2 year
warranty
|
|
Applications
|
|
|
Built
around our high-quality micro-sensor technology, the GX-2003 is
RKI’s smallest personal 5-in-1 gas monitor with a built in sample
pump. Weighing only 11 ounces, it has many features which set the
GX-2003 apart from the competition. For example, it can monitor
the standard confined space gases (LEL combustibles, Oxygen
deficiency, Carbon Monoxide, and Hydrogen Sulfide), however it can
also measure 100% volume combustibles and dynamically display
either % LEL, or % volume with its auto-ranging ability.
The GX-2003’s large LCD display shows all gas readings, battery
level, current time, and will automatically backlight in alarm
conditions. Each channel has 2 adjustable alarm levels in addition
to TWA and STEL readings for Carbon Monoxide and Hydrogen Sulfide.
The TWA and STEL readings are equipped with a lunch-break mode,
which prevent the average readings from being reset when the unit
is turned off. Standard alarm types include vibration, visual, and
audible alarms that can be set to latching or non-latching.
Controlled by a microprocessor, the GX-2003 continuously checks
itself for sensor connections, low battery, circuit trouble, low
flow, and calibration errors. The GX-2003 can interchangeably
operate on either a Ni-Cad pack or 3 AA alkaline batteries. The
batteries are simple to replace requiring no tools to access the
battery compartment. The Ni-Cad battery pack can also receive a
charge directly within the instrument or separately, allowing the
unit to remain in service.
The GX-2003 is equipped with datalogging and a sizable memory
capacity of up to 600 hours of log time. With this much capacity,
the GX-2003 can store 100 alarm events, 100 trouble events, the
last 20 calibrations, and up to 8 alarm trends where data is
collected automatically every 5 seconds for 30 minutes before and
after an alarm occurs. Calibration intervals and reminders are
user adjustable and can be set to either go into alarm or to lock
the user out of normal measurement mode once a calibration period
has expired. Calibrations can be performed on all channels
automatically or individually in single calibration mode. The
GX-2003 is also compatible with the Data Cal 2000 calibration
station for simultaneous, hands-free service of up to 10
instruments. |
|
| Gas Detected |
All Combustible Gases
(Methane as standard) |
% Volume Methane
|
Oxygen
(O2) |
Hydrogen Sulfide
(H2S) |
Carbon Monoxide
(CO) |
| Detection Principle |
Catalytic combustion |
Thermal conductivity |
Galvanic cell |
Electro-chemical
cell |
| Detection Range |
0~100% LEL |
0~100% Vol. |
0~40.0% Vol. |
0~100.0 ppm |
0~500 ppm |
| Sampling Method |
Internal sample
pump, flow rate nominal 0.5 LPM |
| Display |
Digital LCD with 7
segments, auto backlight during alarm |
| Preset Alarms (User Adjustable) |
- 1st alarm 10% LEL
- 2nd alarm 50% LEL
- Over alarm 100% LEL
|
N/A |
- Low alarm 19.5%
- High alarm 23.5%
- Over alarm 40.0%
|
- 1st alarm 10 ppm
- 2nd alarm 30 ppm
- TWA alarm 10 ppm
STEL alarm 15 ppm
- Over alarm 100 ppm
|
1st alarm 25 ppm
2nd alarm 50 ppm
TWA 25 ppm
STEL alarm 200 ppm
Over alarm 500 ppm |
| Alarm Method |
Gas alarms: 1st and 2nd (user adjustable), STEL,
TWA and OVER
Trouble alarms: sensor connection, low battery, low flow,
circuit trouble and calibration error
|
| Types of Alarms |
Gas alarms: flashing lights, two tone buzzer, and
vibration
Trouble alarms: flashing lights, trouble displayed, intermittent
buzzer, and vibration
|
| Special Functions |
Manual back light LCD (Automatic at alarm)
Datalogging (up to 600 hours)
STEL, TWA
Alarm trend data for last 8 alarms
Peak hold
Direct charging in just 90 minutes (Ni-Cad version)
Continuous display of current time
Calibration reminder and lock out
Autocalibration or single calibration
|
| Operating Temp. & Humidity |
-10° C ~ + 40°C (14
to 104°F), 0 to 85% RH, non-condensing |
| Response Time |
Within 30 seconds
(T90) |
| Continuous Operation |
Alkaline battery: 14
hours
Ni-Cad battery: 10 hours (fully charged) |
| Power Source |
Ni-Cad battery pack,
direct or separate charge, or 3 “AA” Alkaline dry cell batteries |
| Safety Design |
Intrinsically safe
design, CSA, C / US approved, CSA file No. 186718-2500028218 |
| Dimensions & Weight |
Approx. 171 (H) x 65
(W) x 39 (D) mm (5.6” H x 2.5”W x 1.5”D), approx. 310 g (11
ounces) |
| Case Material |
Dust & splash
resistant. RFI shielded high impact plastic with protective rubber
boot |
| Controls |
Five buttons: POWER
/ ENTER, DISPLAY, AIR, RESET, SHIFT |
| Warranty |
Two years material
and workmanship |
| Standard Accessories |
Strap
Probe
Rubber boot
10’ hose (some versions)
|
| Optional Accessories |
Charger
Ni-Cad battery
Datalogging software
Calibration kit
Data Cal 2000 calibration station
Sample draw hoses (10’ standard, up to 30’ max.)
|
| Ordering Information |
GX-2003, 5 sensors, LEL / % Vol /
O2 / CO / H2S
GX-2003, 4 sensors, LEL / O2 / H2S / CO
GX-2003, 3 sensors, LEL / O2 / H2S
GX-2003, 3 sensors, LEL / O2 / CO
GX-2003, 2 sensors, LEL / O2
GX-2003, 2 sensors, LEL / 1% Vol / O2 |
*Other gases available for % volume
range.
*Specifications subject to change without notice. |
|